Blind Test Performance of the ESA FTIR-Based Multi-Component Monitoring System for Spacecraft Air Analysis 2001-01-2157
Through the last decade ESA (European Space Agency) have developed a TGM (Trace Gas Monitoring) system for spacecraft air analysis from paper study to a fully operational breadboard [1–9]. The TGM system is a combination of FTIR spectrometry and specially developed analysis techniques. Ongoing work aims at a system flight experiment on e.g. the Space Shuttle early in 2003 [10].
This paper mainly covers the application of the TGM breadboard for a competitive blind testing on unknown multi-gas mixtures arranged by NASA. The TGM HW was applied as is, but a special calibration for the NASA test scenario was made. Four different system suppliers competed in the testing, and the ESA TGM system performed clearly best [11].
This independently defined and supervised testing has confirmed that the ESA TGM system is reliable to perform quasi-real time gas measurements in the concentration ranges required by ESA. It remains to demonstrate that the even lower detection limits required by NASA can be covered through emphasis on improved sensitivity.
Citation: Honne, A., Johansen, I., Stuffler, T., and Tan, G., "Blind Test Performance of the ESA FTIR-Based Multi-Component Monitoring System for Spacecraft Air Analysis," SAE Technical Paper 2001-01-2157, 2001, https://doi.org/10.4271/2001-01-2157. Download Citation
Author(s):
Atle Honne, Ib-Rune Johansen, Timo Stuffler, Gijsbert Tan
Affiliated:
SINTEF
Pages: 18
Event:
31st International Conference On Environmental Systems
ISSN:
0148-7191
e-ISSN:
2688-3627
Related Topics:
Spacecraft
Gases
Suppliers
Reliability
Calibration
SAE MOBILUS
Subscribers can view annotate, and download all of SAE's content.